RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ CSA and CSAE The Cyclosporine method is an in vitro diagnostic test for the quantitative measurement

Recalled May 28, 2014 · FDA recall Z-1647-2014 · Siemens Healthcare Diagnostics

Hazard
A complaint was received of Dimension Vista¿ CSA and CSAE under-recovery of cyclosporine when patient samples are run from the Small Sample Container. The under-recovery that has been observed is 18 to 44%. In March 2014, Siemens recalled due to the complaint received of Dimension Vista¿ CSA and CSAE reporting under-recovery of cyclosporine when patient samples are run from the Small Sample C · FDA Class II
Units
4175
Sold at
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Belgium, Switzerland, Germany, Spain, France, Great Britain, Italy, Saudi Arabia, Australia, and Japan.

Description

A complaint was received of Dimension Vista¿ CSA and CSAE under-recovery of cyclosporine when patient samples are run from the Small Sample Container. The under-recovery that has been observed is 18 to 44%. In March 2014, Siemens recalled due to the complaint received of Dimension Vista¿ CSA and CSAE reporting under-recovery of cyclosporine when patient samples are run from the Small Sample C

Product
Dimension Vista¿ CSA and CSAE The Cyclosporine method is an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CSA) in human whole blood on the Dimension Vista¿ System. Measurements of CSA are used as an aid in the management of heart, liver and kidney transplant patients.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ CSA and CSAE The Cyclosporine method is an in vitro diagnostic test for the quantitative measurement | RecallWatchUSA