Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista Ferritin (FERR) Flex reagent cartridge

Recalled June 28, 2017 · FDA recall Z-2609-2017 · Siemens Healthcare Diagnostics

Hazard
The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere. · FDA Class II
Units
29,266 units
Sold at
Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa

Description

The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.

Product
Dimension Vista Ferritin (FERR) Flex reagent cartridge
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov