Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ HbA1c

Recalled November 5, 2014 · FDA recall Z-0127-2015 · Siemens Healthcare Diagnostics

Hazard
In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t · FDA Class II
Units
20043
Sold at
Worldwide Distribution: US (nationwide) and country of: Canada.

Description

In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t

Product
Dimension Vista¿ HbA1c
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.