Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista HDLC - colorimetric method, lipoproteins - Product Usage: used as an aid in the diagnosis of lipid disor

Recalled March 17, 2021 · FDA recall Z-1188-2021 · Siemens Healthcare Diagnostics

Hazard
Siemens Healthcare Diagnostics Inc. has confirmed the potential for erroneous results due to discolored reagent observed in some wells 5 and 6 of Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridges lot 20062BA. · FDA Class II
Units
4174 devices (Domestic: 2836 devices, Foreign: 1338 devices)
Sold at
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY, and the countries of Austria, Belgium, Canada, Denmark, France, Germany, Italy, Netherlands, Portugal, SA, Slovakia, Slovenia, Spain, Switzerland.

Description

Siemens Healthcare Diagnostics Inc. has confirmed the potential for erroneous results due to discolored reagent observed in some wells 5 and 6 of Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridges lot 20062BA.

Product
Dimension Vista HDLC - colorimetric method, lipoproteins - Product Usage: used as an aid in the diagnosis of lipid disorders (such as diabetes mellitus), various liver and renal diseases, and in the assessment of risk for atherosclerosis and cardiovascular disease.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista HDLC - colorimetric method, lipoproteins - Product Usage: used as an aid in the diagnosis of lipid disor | RecallWatchUSA