Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3.

Recalled September 7, 2022 · FDA recall Z-1648-2022 · Siemens Healthcare Diagnostics

Hazard
Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions. · FDA Class II
Units
129 units US; 159 units OUS
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, Austria, France, ¿¿ Germany, Italy, Switzerland.

Description

Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.

Product
Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number (SMN)/Ref Catalog Number : 10464344/K6456
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.