RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics, Inc recalls Dimension Vista LOCI Estradiol Test Code E2 Catalog Number: K6463 SMN: 10489099 Product Usage: The E2 method is a

Recalled March 30, 2016 · FDA recall Z-1241-2016 · Siemens Healthcare Diagnostics

Hazard
The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results. · FDA Class II
Units
1945 units
Sold at
Worldwide Distribution  US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.

Description

The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.

Product
Dimension Vista LOCI Estradiol Test Code E2 Catalog Number: K6463 SMN: 10489099 Product Usage: The E2 method is an in vitro diagnostic test for the quantitative measurement of estradiol in human serum and plasma on the Dimension Vista¿System. Estradiol measurements are used in the diagnosis and treatment of various hormonal sexual disorders and in assessing placental function in complicated pregnancy.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov