Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.

Recalled August 13, 2025 · FDA recall Z-2272-2025 · Siemens Healthcare Diagnostics

Hazard
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. · FDA Class II
Units
37,554 units
Sold at
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bahrain, Bolivia, Brazil, Canada, Canary Islands, Chile, Croatia, Denmark, Ecuador, Estonia, France, Fren. Polynesia, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Caledonia, Paraguay, Philippines, Poland, Portugal, Qatar, Rep. Korea (S), Romania, Russian Fed., San Marino, Slovakia, Slovenia, South Africa, Spain, Srilanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, Unit.Arab Emir., United Kingdom, Uruguay, Japan.

Description

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.

Product
epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.
Manufacturer(s)
Siemens Healthcare Diagnostics Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov