Home / Recalls / Medical Devices
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi Gastrin
Recalled May 16, 2018 · FDA recall Z-1681-2018 · Siemens Healthcare Diagnostics
Hazard
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin. · FDA Class II
Units
N/A
Sold at
Nationwide and Canada, Mexico, and Thailand
Description
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
Product
IMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi Gastrin
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →