Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Blood Gas Analyzer, Siemens Material Numbers (SMN):10492730, 10696855, 10696857, 10697306 This syste

Recalled September 9, 2015 · FDA recall Z-2479-2015 · Siemens Healthcare Diagnostics

Hazard
When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to include either: duplicate data, missing data, or data from a different patient. · FDA Class II
Units
5241 units
Sold at
Worldwide Distribution - US (nationwide) Distribution and to the countries of : Albania, Algeria, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg, Brazil, Bulgaria,Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany,Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Italy, Japan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mexico, Netherlands. New Zealand, Norway, P.R. China, Peru, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, U.A.E., United Kingdom and Vietnam.

Description

When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to include either: duplicate data, missing data, or data from a different patient.

Product
RAPIDPoint 500 Blood Gas Analyzer, Siemens Material Numbers (SMN):10492730, 10696855, 10696857, 10697306 This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples.
Manufacturer(s)
Siemens Healthcare Diagnostics Inc