RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics Inc recalls Siemens Chek-Stix Combo Pak (Distributed OUS) Catalog Number: 1364E Chek-Stix¿ Positive and Negative Control Strips

Recalled January 16, 2013 · FDA recall Z-0641-2013 · Siemens Healthcare Diagnostics

Hazard
When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F). · FDA Class III
Units
4642
Sold at
Worldwide Distribution - USA (nationwide) and internationally to Canada, Aruba, Brazil, Denmark, India, Mexico, New Zealand, Singapore,Thailand, Bahamas, South Wales, and Hong Kong,

Description

When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25¿C (77¿F).

Product
Siemens Chek-Stix Combo Pak (Distributed OUS) Catalog Number: 1364E Chek-Stix¿ Positive and Negative Control Strips for Urine Chemistry: When used with Siemens reagent strips and tablets for urinalysis, CHEK-STIX¿ positive and negative control strips provide defined results for glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogin, nitrite, and leukocytes
Manufacturer(s)
Siemens Healthcare Diagnostics Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Healthcare Diagnostics Inc recalls Siemens Chek-Stix Combo Pak (Distributed OUS) Catalog Number: 1364E Chek-Stix¿ Positive and Negative Control Strips | RecallWatchUSA