Siemens Healthcare Diagnostics, Inc. recalls Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A clinica

Recalled September 10, 2014 · FDA recall Z-2561-2014 · Siemens Healthcare Diagnostics

Hazard
May not perform as intended under certain conditions, causing the release of results to the Laboratory Information System (LIS) that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS. · FDA Class II
Units
1347
Sold at
US Nationwide Distribution

Description

May not perform as intended under certain conditions, causing the release of results to the Laboratory Information System (LIS) that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.

Product
Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A clinical data management system that assists medical laboratory professionals with preanalytic and post-analytic functions in conjunction with multiple instruments, the laboratory information system (LIS) and Siemens StreamLAB Automation Solutions.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.