Siemens Medical Solutions USA, Inc recalls ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
Recalled April 15, 2026 · FDA recall Z-1653-2026 · Siemens Medical Solutions USA
Hazard
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements. · FDA Class II
Units
U.S. 758, OUS 2077
Sold at
Worldwide - US Nationwide distribution.
Description
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
Product
ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
Manufacturer(s)
Siemens Medical Solutions USA, Inc