RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Medical Solutions USA, Inc recalls Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000

Recalled February 2, 2022 · FDA recall Z-0515-2022 · Siemens Medical Solutions USA

Hazard
Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option · FDA Class II
Units
12 units
Sold at
US Nationwide distribution.

Description

Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option

Product
Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Medical Solutions USA, Inc recalls Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000 | RecallWatchUSA