Siemens Medical Solutions USA, Inc recalls Artis Zee and Zeego fluoroscopic x-ray system

Recalled March 22, 2017 · FDA recall Z-1467-2017 · Siemens Medical Solutions USA

Hazard
A gap in the housing of the wireless foot switch could potentially result in liquids penetrating the interior. Such liquids may include disinfectants, cleaning agents and/ or bodily fluids. In rare cases, the introduction of liquids in the wireless foot switch housing could result in the failure of the foot switch. · FDA Class II
Units
2158 units distributed in US and 6,139 worldwide
Sold at
Distributed Nationwide

Description

A gap in the housing of the wireless foot switch could potentially result in liquids penetrating the interior. Such liquids may include disinfectants, cleaning agents and/ or bodily fluids. In rare cases, the introduction of liquids in the wireless foot switch housing could result in the failure of the foot switch.

Product
Artis Zee and Zeego fluoroscopic x-ray system
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov