RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Medical Solutions USA, Inc recalls Artis zee ceiling, Model Number 10094137

Recalled April 27, 2022 · FDA recall Z-0968-2022 · Siemens Medical Solutions USA

Hazard
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT". · FDA Class II
Units
709 systems
Sold at
US Nationwide.

Description

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Product
Artis zee ceiling, Model Number 10094137
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Medical Solutions USA, Inc recalls Artis zee ceiling, Model Number 10094137 | RecallWatchUSA