Siemens Medical Solutions USA, Inc recalls Artis zee ceiling, Model Number 10094137

Recalled July 31, 2019 · FDA recall Z-2029-2019 · Siemens Medical Solutions USA

Hazard
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure. · FDA Class II
Units
428
Sold at
US Nationwide Distribution

Description

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Product
Artis zee ceiling, Model Number 10094137
Manufacturer(s)
Siemens Medical Solutions USA, Inc