Siemens Medical Solutions USA, Inc recalls Artis zee/ zeego systems; dedicated angiography systems developed for single and biplane diagnostic imaging and interven

Recalled August 26, 2015 · FDA recall Z-2404-2015 · Siemens Medical Solutions USA

Hazard
in case a system error occurs and the system enters the "Bypass Fluoro" mode while the X-ray locking function is active, the only way to exit the X-ray locking function, would be to either resolve the root cause of the system being in "Bypass Fluoro" or to restart the system. · FDA Class II
Units
586
Sold at
Nationwide Distribution

Description

in case a system error occurs and the system enters the "Bypass Fluoro" mode while the X-ray locking function is active, the only way to exit the X-ray locking function, would be to either resolve the root cause of the system being in "Bypass Fluoro" or to restart the system.

Product
Artis zee/ zeego systems; dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov