Siemens Medical Solutions USA, Inc recalls AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single
Recalled August 1, 2018 · FDA recall Z-2525-2018 · Siemens Medical Solutions USA
Hazard
Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure · FDA Class II
Units
6 units US
Sold at
HI , IL, MI, NC, NY, VA, WI
Description
Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure
Product
AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.
Manufacturer(s)
Siemens Medical Solutions USA, Inc