Siemens Medical Solutions USA, Inc recalls AXIOM MULTIX MT - Intended Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO
Recalled July 3, 2024 · FDA recall Z-2181-2024 · Siemens Medical Solutions USA
Hazard
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire · FDA Class II
Units
7 units
Sold at
US Nationwide distribution.
Description
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
Product
AXIOM MULTIX MT - Intended Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO and TOP Systems radiographic devices Material Number: 8395415
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →