Siemens Medical Solutions USA, Inc. recalls Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner.

Recalled May 9, 2018 · FDA recall Z-1597-2018 · Siemens Medical Solutions USA

Hazard
Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect values are being applied. This can lead to a gradient in the image. The severity of the error is directly related to the positioning of the PET QC phantom relative to the center of the field of view. · FDA Class II
Units
42 units
Sold at
Worldwide distribution. US nationwide, Australia, France, India, Japan, and Netherlands.

Description

Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect values are being applied. This can lead to a gradient in the image. The severity of the error is directly related to the positioning of the PET QC phantom relative to the center of the field of view.

Product
Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner.
Manufacturer(s)
Siemens Medical Solutions USA, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov