Home / Recalls / Medical Devices
Siemens Medical Solutions USA Inc. recalls BIOGRAPH mCT S(20) - 4R, MATERIAL NUMBER 10528958 Product Usage: The Siemens Biograph mCT systems are combined X-Ray
Recalled June 22, 2016 · FDA recall Z-2000-2016 · Siemens Medical Solutions USA
Hazard
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service. · FDA Class II
Units
1 unit
Sold at
Worldwide Distribution - US Nationwide and the countries of China and Malaysia.
Description
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
Product
BIOGRAPH mCT S(20) - 4R, MATERIAL NUMBER 10528958 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Manufacturer(s)
Siemens Medical Solutions USA Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →