Siemens Medical Solutions USA Inc. recalls Biograph mCT-S(40) 3R, MATERIAL NUMBER 10248668 The Siemens Biograph TruePoint systems are combined X-Ray Computed To

Recalled June 29, 2016 · FDA recall Z-2030-2016 · Siemens Medical Solutions USA

Hazard
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration. · FDA Class II
Units
47 units
Sold at
Worldwide Distribution - US Nationwide

Description

Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.

Product
Biograph mCT-S(40) 3R, MATERIAL NUMBER 10248668 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Manufacturer(s)
Siemens Medical Solutions USA Inc.