Home / Recalls / Medical Devices
Siemens Medical Solutions USA, Inc recalls Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Al
Recalled September 2, 2020 · FDA recall Z-2848-2020 · Siemens Medical Solutions USA
Hazard
Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality. · FDA Class II
Units
US: 282 units; OUS: 1363
Sold at
Nationwide
Description
Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.
Product
Cios Alpha (VA20) - Model: 10308191 Cios Alpha (VA30) - Model: 11105200 Cios Spin (VA30) - Model: 10308194 Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →