Home / Recalls / Medical Devices
Siemens Medical Solutions USA, Inc recalls MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking st
Recalled May 11, 2016 · FDA recall Z-1598-2016 · Siemens Medical Solutions USA
Hazard
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet. · FDA Class II
Units
8 systems
Sold at
US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.
Description
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
Product
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →