Siemens Medical Solutions USA, Inc recalls MAGNETOM Lumina (DE). Model Number: 11344916.
Recalled October 8, 2025 · FDA recall Z-2658-2025 · Siemens Medical Solutions USA
Hazard
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room. · FDA Class I
Units
130 units
Sold at
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
Description
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Product
MAGNETOM Lumina (DE). Model Number: 11344916.