RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Medical Solutions USA, Inc recalls MAGNETOM Skyra magnetic resonance diagnostic devices Model # 10432915

Recalled August 28, 2019 · FDA recall Z-2343-2019 · Siemens Medical Solutions USA

Hazard
Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly. · FDA Class II
Units
44
Sold at
AL AR AZ CA CO CT DC DE FL GA GU IA ID IL IN KS KY LA MA MD ME MI MN MO MT NC NE NH NJ NY OH OK OR PA SC TN TX VA WA WI WV WY

Description

Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.

Product
MAGNETOM Skyra magnetic resonance diagnostic devices Model # 10432915
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov