Siemens Medical Solutions USA, Inc recalls MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHE

Recalled September 22, 2021 · FDA recall Z-2447-2021 · Siemens Medical Solutions USA

Hazard
Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS). · FDA Class II
Units
2 systems
Sold at
US Nationwide distribution.

Description

Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).

Product
MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815
Manufacturer(s)
Siemens Medical Solutions USA, Inc