Siemens Medical Solutions USA, Inc recalls MULTIX FUSION. The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. M

Recalled April 30, 2014 · FDA recall Z-1505-2014 · Siemens Medical Solutions USA

Hazard
It was observed that the corrugated hose holder above the 3D rail on the Multix Fusion system became loose. According to the Installation Instructions the corrugated hose holder shall be afixed with transversal carriage, with Loctite 243 and Torque 10nm during system installation. Siemens initiated the update instruction to verify and secure the hose holder with Loctite 243. A PCB D802 Rev. 00 b · FDA Class II
Units
12
Sold at
Distributed in the states of AK, PA, KY, OH, WI, MI, FL, and NY.

Description

It was observed that the corrugated hose holder above the 3D rail on the Multix Fusion system became loose. According to the Installation Instructions the corrugated hose holder shall be afixed with transversal carriage, with Loctite 243 and Torque 10nm during system installation. Siemens initiated the update instruction to verify and secure the hose holder with Loctite 243. A PCB D802 Rev. 00 b

Product
MULTIX FUSION. The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov