RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Medical Solutions USA, Inc recalls ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIMUS HI, MEVATRON M2/PRIMUS Mid-Energy,

Recalled November 4, 2015 · FDA recall Z-0168-2016 · Siemens Medical Solutions USA

Hazard
potential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option. · FDA Class II
Units
158
Sold at
Nationwide Distribution

Description

potential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option.

Product
ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIMUS HI, MEVATRON M2/PRIMUS Mid-Energy, ARTISTE; intended use is to deliver x-ray radiation for therapeutic treatment of cancer.
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Medical Solutions USA, Inc recalls ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIMUS HI, MEVATRON M2/PRIMUS Mid-Energy, | RecallWatchUSA