Home / Recalls / Medical Devices
Siemens Medical Solutions USA, Inc recalls Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system
Recalled July 24, 2013 · FDA recall Z-1750-2013 · Siemens Medical Solutions USA
Hazard
During the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may fail. · FDA Class II
Units
23
Sold at
United States Nationwide Distribution - including the states of: CA, CO, FL, IA, IL, IN, MA, MD, MI, NC, NY, PA, TX, VA and WA and Puerto Rico.
Description
During the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may fail.
Product
Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →