Siemens Medical Solutions USA, Inc recalls SIEMENS Healthineers Luminos Agile Max Model: 10762472 - Product Usage: is a device intended to visualize anatomical s
Recalled August 12, 2020 · FDA recall Z-2756-2020 · Siemens Medical Solutions USA
Hazard
Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis · FDA Class II
Units
477 units
Sold at
US Nationwide distribution.
Description
Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis
Product
SIEMENS Healthineers Luminos Agile Max Model: 10762472 - Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →