Home / Recalls / Medical Devices
Siemens Medical Solutions USA, Inc recalls Siemens Linear Accelerator (LINAC) models. Product Usage: deliver X-ray radiation for therapeutic treatment of canc
Recalled June 4, 2014 · FDA recall Z-1652-2014 · Siemens Medical Solutions USA
Hazard
Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models. · FDA Class II
Units
9
Sold at
US Nationwide Distribution in the states of AL, ME, PA, IL, and CA.
Description
Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models.
Product
Siemens Linear Accelerator (LINAC) models. Product Usage: deliver X-ray radiation for therapeutic treatment of cancer
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →