Siemens Medical Solutions USA, Inc recalls Siemens Mobilett Mira. Mobile X-Ray System.

Recalled September 12, 2012 · FDA recall Z-2322-2012 · Siemens Medical Solutions USA

Hazard
During regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system. During normal clinical operation the switch on the motor drive handle of the unit may become loose and slide horizontally to the right and become trapped under the cover. This may cause the system to continuously move. In response, Siemens has issued Update Instruction XP038/12/S to r · FDA Class II
Units
20
Sold at
Nationwide Distribution- including the states of GA, ID, IL, IA, KS, NC, OH, SD, TX, and WI.

Description

During regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system. During normal clinical operation the switch on the motor drive handle of the unit may become loose and slide horizontally to the right and become trapped under the cover. This may cause the system to continuously move. In response, Siemens has issued Update Instruction XP038/12/S to r

Product
Siemens Mobilett Mira. Mobile X-Ray System.
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov