Siemens Medical Solutions USA, Inc recalls Siemens syngo RT Dosimetrist 2.7 system Product Usage: The intended use of syngo Suite for Oncology Systems is as an

Recalled July 23, 2014 · FDA recall Z-2075-2014 · Siemens Medical Solutions USA

Hazard
An update to the Virtual Simulation software of the syngo RT Dosimetrist is necessary to address a safety issue. In special scenarios within the Structure Operations function the system did not store the selected name from the list of predefined names. After reloading of the data, this structure name had been changed by the system. Additionally, a performance issue was addressed where import of DI · FDA Class II
Units
7
Sold at
US Nationwide Distribution in the states of NH, TN, OH, VA, and TX.

Description

An update to the Virtual Simulation software of the syngo RT Dosimetrist is necessary to address a safety issue. In special scenarios within the Structure Operations function the system did not store the selected name from the list of predefined names. After reloading of the data, this structure name had been changed by the system. Additionally, a performance issue was addressed where import of DI

Product
Siemens syngo RT Dosimetrist 2.7 system Product Usage: The intended use of syngo Suite for Oncology Systems is as an accessory to the linear accelerator system to aid and support in the planning and delivery of x-ray radiation for the therapeutic treatment of cancer.
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov