RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Medical Solutions USA, Inc recalls SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system

Recalled November 27, 2019 · FDA recall Z-0265-2020 · Siemens Medical Solutions USA

Hazard
SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures · FDA Class II
Units
14 US
Sold at
US in the states of: GA, IN, MO, MS, NJ, OH, PA

Description

SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures

Product
SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov