Siemens Medical Solutions USA, Inc recalls SOMATOM Perspective, Computed tomography x-ray system
Recalled January 4, 2017 · FDA recall Z-0921-2017 · Siemens Medical Solutions USA
Hazard
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems. · FDA Class II
Units
13 systems
Sold at
Distributed to: IL IL GA SC SD OK NY MI IN MD AR
Description
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
Product
SOMATOM Perspective, Computed tomography x-ray system
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →