Siemens Medical Solutions USA, Inc recalls SOMATOM Perspective; the intended use of computed tomography is to produce cross-sectional images of the body by compute

Recalled June 3, 2015 · FDA recall Z-1646-2015 · Siemens Medical Solutions USA

Hazard
Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks. · FDA Class II
Units
75
Sold at
Nationwide Distribution including Puerto Rico.

Description

Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.

Product
SOMATOM Perspective; the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.
Manufacturer(s)
Siemens Medical Solutions USA, Inc