Siemens Medical Solutions USA, Inc recalls SOMATOM Sensation 40 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transm

Recalled May 9, 2018 · FDA recall Z-1458-2018 · Siemens Medical Solutions USA

Hazard
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure · FDA Class II
Units
1301
Sold at
USA (nationally) Distribution.

Description

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Product
SOMATOM Sensation 40 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
Manufacturer(s)
Siemens Medical Solutions USA, Inc