Home / Recalls / Medical Devices
Siemens Medical Solutions USA, Inc recalls Syngo Imaging XS is a Picture Archiving and Communication System (PACS)
Recalled December 30, 2015 · FDA recall Z-0550-2016 · Siemens Medical Solutions USA
Hazard
For the Syngo Workflow SLR System with Software Ver: VB10C: Printouts may be printed in incorrect anatomical size when using syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo Imaging XS printer configuration file. · FDA Class II
Units
42
Sold at
AL ; CA ; CO ; CT ; FL ; GA ; IA ; IL ; IN ; KS ; MD ; MI ; MO ; NC ; NE ; NJ ; NV ; NY ; OH ; PA ; TN ; TX ; WI ;
Description
For the Syngo Workflow SLR System with Software Ver: VB10C: Printouts may be printed in incorrect anatomical size when using syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo Imaging XS printer configuration file.
Product
Syngo Imaging XS is a Picture Archiving and Communication System (PACS)
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →