Siemens Medical Solutions USA, Inc recalls syngo.via RT Image Suite with syngo.via VB30, VB40, VB50 and VB60- Intended for Radiation Therapy Treatment Material

Recalled September 22, 2021 · FDA recall Z-2445-2021 · Siemens Medical Solutions USA

Hazard
Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS). · FDA Class II
Units
126 units
Sold at
US Nationwide distribution.

Description

Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).

Product
syngo.via RT Image Suite with syngo.via VB30, VB40, VB50 and VB60- Intended for Radiation Therapy Treatment Material Number: 10496180
Manufacturer(s)
Siemens Medical Solutions USA, Inc