Siemens Medical Solutions USA, Inc recalls syngo.via syngo.CT Cardiac Planning, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Ca
Recalled April 17, 2019 · FDA recall Z-1113-2019 · Siemens Medical Solutions USA
Hazard
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm. · FDA Class II
Units
112 total
Sold at
US Nationwide Distribution in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, IA, IL, KS, KY, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
Description
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.
Product
syngo.via syngo.CT Cardiac Planning, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Cardiac Function are image analysis software packages for evaluating cardiac CT angiography (CTA) volume sets. The software packages are designed to support the physician in determining the functional parameters of the left and right ventricles, confirming the presence or absence of physicianidentified myocardial enhancement defects and evaluation documentation and follow-up of any such finding.
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →