Siemens Medical Solutions USA, Inc recalls Syngo.Via VB20A model 1049610
Recalled May 16, 2018 · FDA recall Z-1778-2018 · Siemens Medical Solutions USA
Hazard
Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values. · FDA Class II
Units
N/A
Sold at
Nationwide
Description
Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values.
Product
Syngo.Via VB20A model 1049610
Manufacturer(s)
Siemens Medical Solutions USA, Inc