Home / Recalls / Medical Devices
Siemens Medical Solutions USA, Inc recalls Syngo.x picture archiving and communication system Syngo.x is a software solution intended to be used for viewing, ma
Recalled September 28, 2016 · FDA recall Z-2852-2016 · Siemens Medical Solutions USA
Hazard
Software error. Incorrect values for the volume calculation from a freehand VOI at the customer site. In volume calculations of prostate as well as in volume calculations of liver were too high. · FDA Class II
Units
338 systems
Sold at
Distributed Nationwide
Description
Software error. Incorrect values for the volume calculation from a freehand VOI at the customer site. In volume calculations of prostate as well as in volume calculations of liver were too high.
Product
Syngo.x picture archiving and communication system Syngo.x is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images.
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →