RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Medical Solutions USA, Inc recalls UROSKOP Omnia. Model Number: 10094910

Recalled June 25, 2025 · FDA recall Z-1981-2025 · Siemens Medical Solutions USA

Hazard
A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire. · FDA Class II
Units
95 units
Sold at
Nationwide distribution.

Description

A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.

Product
UROSKOP Omnia. Model Number: 10094910
Manufacturer(s)
Siemens Medical Solutions USA, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov