Siemens Medical Solutions USA, Inc recalls Ysio X.pree, Model Number: 11107464 - Product Usage: A Digital X-Ray System generate Xray images from the whole body in
Recalled April 14, 2021 · FDA recall Z-1346-2021 · Siemens Medical Solutions USA
Hazard
A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted · FDA Class II
Units
12 units
Sold at
US Nationwide distribution in the states of FL, MN, OH, PA, TX.
Description
A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted
Product
Ysio X.pree, Model Number: 11107464 - Product Usage: A Digital X-Ray System generate Xray images from the whole body including the skull, chest, abdomen, and extremities.
Manufacturer(s)
Siemens Medical Solutions USA, Inc