Home / Recalls / Medical Devices
Siemens Medical Solutions USA, Inc recalls ZE for Vertix Solitaire - The AXIOM Vertix Solitaire is a component and has been specially designed for examining emerge
Recalled July 3, 2024 · FDA recall Z-2188-2024 · Siemens Medical Solutions USA
Hazard
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire · FDA Class II
Units
2 units
Sold at
US Nationwide distribution.
Description
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
Product
ZE for Vertix Solitaire - The AXIOM Vertix Solitaire is a component and has been specially designed for examining emergency and accident patients as well as for use in outpatient departments Material Number: 5901900
Manufacturer(s)
Siemens Medical Solutions USA, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →