Home / Recalls / Medical Devices
Smith & Nephew, Inc. recalls D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD
Recalled April 18, 2018 · FDA recall Z-1317-2018 · Smith & Nephew
Hazard
One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch. · FDA Class II
Units
59 units
Sold at
US, Puerto Rico, Malaysia
Description
One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.
Product
D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R
Manufacturer(s)
Smith & Nephew, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →