RecallWatchUSA

Home / Recalls / Medical Devices

Smith & Nephew, Inc. recalls smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.

Recalled May 30, 2018 · FDA recall Z-1921-2018 · Smith & Nephew

Hazard
Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm. · FDA Class II
Units
13 units
Sold at
US nationwide distribution.

Description

Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.

Product
smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.
Manufacturer(s)
Smith & Nephew, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Smith & Nephew, Inc. recalls smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis. | RecallWatchUSA