RecallWatchUSA

Home / Recalls / Medical Devices

Smith & Nephew, Inc. recalls smith&nephew EVOS CONDYLAR MEDIAL DISTAL FEMUR PLATE, R, 5H, 115MM, REF 72573202

Recalled December 22, 2021 · FDA recall Z-0366-2022 · Smith & Nephew

Hazard
Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System. · FDA Class II
Units
41 units
Sold at
US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.

Description

Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.

Product
smith&nephew EVOS CONDYLAR MEDIAL DISTAL FEMUR PLATE, R, 5H, 115MM, REF 72573202
Manufacturer(s)
Smith & Nephew, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Smith & Nephew, Inc. recalls smith&nephew EVOS CONDYLAR MEDIAL DISTAL FEMUR PLATE, R, 5H, 115MM, REF 72573202 | RecallWatchUSA