Home / Recalls / Medical Devices
Smith & Nephew, Inc. recalls smith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201
Recalled December 22, 2021 · FDA recall Z-0364-2022 · Smith & Nephew
Hazard
Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System. · FDA Class II
Units
35 units
Sold at
US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.
Description
Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.
Product
smith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201
Manufacturer(s)
Smith & Nephew, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →